Myocardial single-photon emission computed tomography (SPECT) is considered as the gold standard for the diagnosis of coronary artery disease. Developed in the 1980s with rotating Anger gammacameras, this technique could be dramatically enhanced by new imaging systems working with semiconductor detectors and which performances are clearly enhanced. Two semiconductor cameras, dedicated to nuclear cardiology and equipped with Cadmium Zinc Telluride detectors, have been recently commercialized: the Discovery NM- 530c (General Electric) and the DSPECT (Spectrum Dynamics). The performances of these CZT cameras were compared: 1) by a comprehensive analysis of phantom and human SPECT images considered as normal and 2) with the parameters commonly recommended for SPECT recording and reconstruction. The results show the superiority of the CZT cameras in terms of detection sensitivity, spatial resolution and contrast-to-noise ratio, compared to conventional Anger cameras. These properties might lead to dramatically reduce acquisition times and/or the injected activities. However, the limits of these new CZT cameras, as well as the mechanism of certain artefacts, remain poorly known. This knowledge could be enhanced by a numerical modeling of the DSPECT camera, and this might also help to optimize acquisition and reconstruction parameters. We developed a simulator where the geometry of the detectors of the DSPECT camera and their energy response were modeled in the GATE platform. In order to validate this simulator, actually recorded data were compared with simulated data through three performance parameters: detection sensitivity, spatial resolution and energy resolution. Results were in agreement between simulated and actually recorded data. This observation validates the DSPECT simulator and opens the door to further studies planed to optimize the recorded and reconstruction processes, especially for complex protocols such as simultaneous dual-radionuclide acquisition and kinetics firstpass studies.