An evaluation of CT-scan to locate the femoral head centre and its implication for hip surgeons

The aim of this preliminary study was to determine the accuracy of CT-scan to locate the femoral head centre. METHODS: Eleven dried femurs were included for study. Three techniques were compared to determine femoral head centre (FHC) location: CT-scan, Motion Analysis and Faro-Arm. Markers were stuck on each femur to create a system of coordinates. Femurs lied on their posterior parts (bicondylar plane). Several points around the femoral head were palpated (Motion Analysis and Faro-Arm) or determined (Amira software for CT-scans). By a least-square regression method, the FHC location in 3D was defined for each technique. RESULTS: The results of the FHC location determined by the CT-scan technique were compared with those measured by the faro-arm and the Motion Analysis techniques. The coordinates (X, Y, Z) of the FHC were compared between the three methods, and no statistical difference was found (p = 0.99). In a 3D plot, this gave a mean difference of 1.3 mm. The mean radius of the femoral head was of 22.5 mm (p = 0.6). CONCLUSIONS: CT-scan is as accurate and reliable as gold-standard techniques (motion and faro-arm). Locating FHC before and after hip arthroplasty would allow hip surgeons to determine and compare 3D orientation of the upper-end of femur: offset, height and anteversion.

Data and Resources

Additional Info

Field Value
Source ISSN: 0930-1038
Author Viste, Anthony, Trouillet, Franck, Testa, Rodolphe, Cheze, Laurence, Desmarchelier, Romain, Fessy, Michel-Henri
Maintainer CCSD
Last Updated May 10, 2026, 00:39 (UTC)
Created May 10, 2026, 00:39 (UTC)
Identifier hal-00854135
Language en
contributor Faculté de Médecine Lyon-Sud ; FACULTE DE LYON
creator Viste, Anthony
date 2014-01-01T00:00:00
harvest_object_id 45d95eb4-678f-4071-a66c-0d1852349286
harvest_source_id 3374d638-d20b-4672-ba96-a23232d55657
harvest_source_title test moissonnage SELUNE
metadata_modified 2024-12-03T00:00:00
relation info:eu-repo/semantics/altIdentifier/doi/10.1007/s00276-013-1172-7
set_spec type:ART